TappingMode AFM
Engaging The Microscope
114 MultiMode SPM Instruction Manual Rev. B
Figure 7.3b Suggested Other Controls Settings During TappingMode Setup
2. Move the probe to the area of interest using the X-Y translation knobs on the MultiMode
head.
3. Use the meters to verify that the vertical deflection is between -1 and +1, the RMS amplitude
(topmost meter) is 1-2V, and the sum voltage is greater than 1V.
4. The operator should now be ready to engage. Go to the upper Realtime menu bar and click
on Motor followed by Engage. You should now see a pre-engage check. If for any reason
the engage aborts because the tip is still too far away from the surface, use the coarse
adjustment screws and/or the Tip Down motor control switch to bring the tip and surface
closer together. If all goes well after re-engaging, a well-formed image will begin to appear
on the display monitor.
If it becomes necessary to image another part of the sample, execute a Withdraw command before
moving the X-Y translation stage; otherwise, the tip may be damaged. After the tip is engaged, the
control panel values can be readjusted to provide the desired scan parameters. Refer to Section 7.5
below for scan parameter optimization.
007