Electric Force (EFM) Imaging
Electric Field Gradient Detection—Preparation
254 MultiMode SPM Instruction Manual Rev. B
External Voltage Source Applied to the Tip
In some cases, it may be advantageous to use voltages greater than 12 VDC, or to use a pulsed
power supply. If an external source of voltage is to be applied to the tip, congure jumpers as
shown in Figure 14.3j.
Note: In all congurations which apply voltage to the tip, an E-eld cantilever holder
is required. Contact Veeco for more information.
Figure 14.3j Jumper conguration for applying external voltage to tip
(for systems with the Basic Extender Module).
A current-limiting resistor (e.g., 10–100M) should be placed in series with the external voltage
supply as shown to protect the tip and sample from damage. Current-limited power supplies may
also be used. Voltage leads should be connected to pins on the header using soldered, push-on
connectors. Do not solder leads directly to the header pins, as the heat may cause damage and/or
make jumpering the pins difcult.
Sample
Tip
+
V
Ground
Analog 2
Unused
Auxiliary D (to NanoScope III controller)
To AFM Tip
Indicates jumpers
Unused
STM
Piezo Cap
(-)(+)
External Voltage
Source
>10 M