Force Imaging
Force Calibration (TappingMode)
202 MultiMode SPM Instruction Manual Rev. B
5. Set the Main Controls and Channel 1 panel parameters to the settings shown in Figure
11.5b.
Note: The Sensitivity value shown in Figure 11.5b may differ from yours.
Figure 11.5b TappingMode Force Plot Parameter Settings (Force Calibrate)
6. Set the Data type parameter to Amplitude under the Channel 1 panel.
7. Adjust the Z scan start parameter to obtain a satisfactory force plot using the left-right arrow
keys.