Force Imaging
Force Calibration Control Panels and Menus
188 MultiMode SPM Instruction Manual Rev. B
11.3.5 Scan Mode Panel (Advanced Mode Only)
Trigger Mode
The Scan Mode panel allows you to use various triggers when obtaining Force Plot and Force
Volume plots. The idea of triggering simple: it limits the total amount of force exerted by the tip
upon the sample. Depending upon which trigger you use and how it is set, you may operate the
trigger independent of drift (Relative) or at some arbitrarily xed point (Absolute).
Figure 11.3b Absolute and Relative Triggers
The plots in Figure 11.3b show the effect of drift on each of the two trigger types. The plot series
shown on the right side of Figure 11.3b utilizes a Relative trigger and maintains force at a constant
level dened by the Trig threshold parameter. The Relative trigger threshold is relative to the data
at the right of the plot where the cantilever is typically undeected. As illustrated, the total force is
maintained, even as the system drifts.
The plot series on the left side of Figure 11.3b utilizes an Absolute trigger. Notice that as the
system drifts (plot moves slowly downward), total force increases. Drift may be due to mechanical
causes, or due to thermal effects on the cantilever. An Absolute trigger permits the total force to be
set using Setpoint values.
Trigger Channel
This parameter denes the channel to watch to determine when to reverse the ramp.
Total Force
Vista de pagina 207