UNIMO Technology UDR-204 Manual de instrucciones Pagina 341

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 358
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 340
Calibration, Maintenance, Troubleshooting and Warranty
TappingMode AFM Troubleshooting
Rev. B MultiMode SPM Instruction Manual 321
15.12.3 Rings around features on the surface
This effect might also be described as the image looking as though it is half submerged beneath
water (see Figure 15.12b). This is caused by operating with a drive frequency too close to cantilever
resonance.
Use the arrow keys to increment the drive frequency a little lower. Do this while watching the
Realtime scan. Be aware that the RMS voltage might also reduce. Repeat the steps in the “lines
across the image” troubleshooting description to adjust the RMS voltage.
Figure 15.12b Rings During High Frequency Operation
023
024
With Rings Without Rings
Vista de pagina 340
1 2 ... 336 337 338 339 340 341 342 343 344 345 346 ... 357 358

Comentarios a estos manuales

Sin comentarios