Scanning Tunneling Microscopy (STM)
Low-Current STM
Rev. B MultiMode SPM Instruction Manual 159
MultiMode SPM
IMPORTANT! If you are installing on a MultiMode SPM equipped with the Basic Extender
Module, you must install the Picoamp Boost box with the jumpers inside set for extended. A label
on the outside of the Boost box indicates whether it was set for standard (Model MMSTMLC) or
Extender (Model MMSTMLCE) at the factory. If you are not certain which base your MultiMode
has, contact Veeco for guidance.
1. Turn off all power to the SPM controller. Disconnect and remove the MultiMode head.
2. Connect the picoamp booster box (male receptacle) to the controller using the 37-pin-to-25-
pin connector cable. If an Basic Extender Module is being used with the system, insert it
between the controller and picoamp booster as diagrammed below in this section.
3. Connect the picoamp booster to the SPM base using the 25-pin-to-25-pin cable connector.
4. Connect the low-current converter head to the MultiMode’s support ring using the micro-D,
15-pin connector cable. Set switches on the picoamp booster to the desired current level and
bandwidth.
5. Before repowering the system, double-check all cable connections. If atomic-level scans are
intended, verify the cable between the SPM and picoamp booster box is unstrained (a taut
cable will transmit vibrations). The cable should rest loosely. Before loading samples and
running the SPM, attach the ESD wristband to prevent sudden electrical discharges.
6. Verify that the current sensitivity settings in the software are set to the appropriate value:
IMPORTANT: The default calibration value of standard STM preamplier electronics is 10 nA/V.
This value must be changed to either 0.1 or 0.01nA/V for operation of the low-current head with
gain factors of either 10
10
and 10
11
respectively. To changes this value, click on Realtime >
Microscope > Calibrate > Z. The Current Sensitivity parameter is on the Z Calibration panel
and can be changed by simply entering the new value.
Controller
Basic Extender
Module (if equipped)
MultiMode
Picoamp Boost Box
37-pin-to-37-pin cable
25-pin-to-25-pin cable
Low-current
Converter Head
Base
37-pin-to-25-pin cable