Magnetic Force (MFM) Imaging
Software Setup Configuration (Basic, Quadrex or NSIV)
234 MultiMode SPM Instruction Manual Rev. B
13.4 Software Setup Configuration (Basic, Quadrex or NSIV)
1. Select di > Microscope Select to display the Microscope Select dialog box (see Figure
13.4a).
Figure 13.4a Microscope Select Dialog Box
2. Click the Edit button to open the Equipment dialog box.
3. Select the appropriate Controller.
4. Select the appropriate Extender.
Note: This step is not necessary for NanoScope IV.
5. Click the Ok button when complete.
13.5 Troubleshooting Suggestions
13.5.1 MFM Image Verification
The procedure described above should produce a good magnetic force gradient image of the
videotape sample. If there is a problem, check that the Interleave Mode is set to Lift, that
Interleave is Enabled and that the Scan Line is set to Interleave. Check also that the Interleave
values of Drive Amplitude and Drive Frequency are initially set equal to the main Scan Controls
values.