UNIMO Technology UDR-204 Manual de instrucciones Pagina 297

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 358
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 296
Calibration, Maintenance, Troubleshooting and Warranty
SPM Calibration Overview
Rev. B MultiMode SPM Instruction Manual 277
Figure 15.1a Scanner Crystal Voltage and Photodiode Voltage
The Microscope > Calibrate > Scanner function displays the Scanner Calibration dialog box,
allowing users to enter the sensitivity of their scanner’s X-Y axes. Here, sensitivity is measured in
terms of lateral displacement for a given voltage (nm/V).
In addition, the NanoScope software employs various derating and coupling parameters to model
scanners’ nonlinear characteristics. By precisely determining points along the scanner’s sensitivity
curve, then applying a rigorous mathematical model, full-range measuring capabilities can be
achieved with greater than 1% accuracy.
-220 VDC 0 VDC +220 VDC
Laser
Photodiode
Scanner
Cantilever
-220 0 +220
Scanner Voltage
Photodiode Voltage
-3.0
0
+3.0
Detector Sensitivity
array
Photodiode voltage
Vista de pagina 296
1 2 ... 292 293 294 295 296 297 298 299 300 301 302 ... 357 358

Comentarios a estos manuales

Sin comentarios