Calibration, Maintenance, Troubleshooting and Warranty
Quick Guide to Piezo Tube Calibration
Rev. B MultiMode SPM Instruction Manual 309
Do not adjust the derating parameters for atomic-scale imaging, including:
•X fast derate
•X slow derate
•Y fast derate
•Y slow derate
Retracted offset der
Extended offset der
As referenced in Section 15.7.6, you must calibrate the sensitivity parameters with the Scan angle
set at both 0 degrees and at 90 degrees.
7. Complete the Z-axis calibration using a silicon calibration reference. Refer to Section 15.8
for instructions on calibrating the Z-axis.
15.10 Quick Guide to Piezo Tube Calibration
15.10.1 Linearize Scanner
This calibration is not necessary during routine maintenance.
1. Engage in Contact AFM with these settings:
Scan size = 440V
Scan rate = 2.44Hz
Scan angle =
2. Rotate the sample so that it is square with respect to the X-Y scan motion.
3. Adjust Orthogonality to make the scan square.
4. Adjust the Fast Mag0 to make the first (1/3) and the last (3/3) of the image have the same
spacing of pits. Typical values are ~1.0 - 2.0.
5. Adjust the Fast Arg to make the middle (2/3) of the image equal to the first (1/3) and last (3/
3) portions of the image. Typical values are ~3.0 - 4.0.
6. Adjust the Slow mag0 and Slow arg to make the slow axis linear. Follow the same procedure
as with the fast axis.